Design and calibration of an optical micro motion sensing system for micromanipulation tasks
Conference proceeding

Design and calibration of an optical micro motion sensing system for micromanipulation tasks

T. L. Win, U. X. Tan, C. Y. Shee, W. T. Ang and IEEE
PROCEEDINGS OF THE 2007 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION, VOLS 1-10, pp.3383-3388
IEEE International Conference on Robotics and Automation ICRA
01/01/2007

Abstract

Automation & Control Systems Robotics Science & Technology Technology
An optical sensing system has been developed using a pair of orthogonally placed position sensitive detectors (PSD) to track 3D displacement of a microsurgical instrument tip in real-time. An infrared (IR) diode is used to illuminate the workspace. A hall is attached to the tip of an intraocular shaft to reflect IR rays onto the PSDs. Instrument tip position is then calculated from the centroid positions of reflected IR light on the respective PSDs. The system can be used to assess the accuracy of hand-held microsurgical instruments and operator performance in micromanipulation tasks, such as microsurgeries. In order to eliminate inherent nonlinearity of the PSDs and lenses, calibration is performed using a feedforward neural network. After calibration, percentage RMS error is reduced from about 5.46 % to about 0.16%. The system RMS noise is about 0.7 mu m. The sampling rate of the system is 250 Hz.