Abstract
Uroflowmetry remains an important tool for the assessment of patients with lower urinary tract symptoms (LUTS), but accuracy can be limited by within-subject variation of urinary flow rates. Voiding acoustics appear to correlate well with conventional uroflowmetry and show promise as a convenient home-based alternative for the monitoring of urinary flows.
To evaluate the ability of a sound-based deep learning algorithm (Audioflow) to predict uroflowmetry parameters and identify abnormal urinary flow patterns.
In this prospective open-label study, 534 male participants recruited at Singapore General Hospital between December 1, 2017 and July 1, 2019 voided into a uroflowmetry machine, and voiding acoustics were recorded using a smartphone in close proximity. The Audioflow algorithm consisted of two models-the first model for the prediction of flow parameters including maximum flow rate (Q
), average flow rate (Q
), and voided volume (VV) was trained and validated using leave-one-out cross-validation procedures; the second model for discrimination of normal and abnormal urinary flows was trained based on a reference standard created by three senior urologists.
Lin's correlation coefficient was used to evaluate the agreement between Audioflow predictions and conventional uroflowmetry for Q
, Q
, and VV. Accuracy of the Audioflow algorithm in the identification of abnormal urinary flows was assessed with sensitivity analyses and the area under the receiver operating curve (AUC); this algorithm was compared with an external panel of graders comprising six urology residents/general practitioners who separately graded flow patterns in the validation dataset.
A total of 331 patients were included for analysis. Agreement between Audioflow and conventional uroflowmetry for Q
, Q
, and VV was 0.77 (95% confidence interval [CI], 0.72-0.80), 0.85 (95% CI, 0.82-0.88) and 0.84 (95% CI, 0.80-0.87), respectively. For the identification of abnormal flows, Audioflow achieved a high rate of agreement of 83.8% (95% CI, 77.5-90.1%) with the reference standard, and was comparable with an external panel of six residents/general practitioners. AUC was 0.892 (95% CI, 0.834-0.951), with high sensitivity of 87.3% (95% CI, 76.8-93.7%) and specificity of 77.5% (95% CI, 61.1-88.6%).
The results of this study suggest that a deep learning algorithm can predict uroflowmetry parameters and identify abnormal urinary voids based on voiding sounds, and shows promise as a simple home-based alternative to uroflowmetry in the management of patients with LUTS.
In this study, we trained a deep learning-based algorithm to measure urinary flow rates and identify abnormal flow patterns based on voiding sounds. This may provide a convenient, home-based alternative to conventional uroflowmetry for the assessment and monitoring of patients with lower urinary tract symptoms.